MICADO MSMThe MSM is located inside the MICADO cryostat, where the operating temperature is 80 K. It will allow to switch between the 4 operational modes of MICADO. The MSM assembly (see Figure) consists of a rotating platform with a diameter of ~1.3 m and its support structure.The rotating platform hosts three optical modules: A cut in the platform provide the clearance for the High Resolution Imager mode. To switch between the 4 operational modes, the platform can be rotated in order to locate the respective optical module in between the HRI fixed optics (in green), along the instrument optical path. The LRI module uses two flat mirrors blocking some of the fixed optical components to change the plate scale on the detector mosaic. The SPE consists of two echelle like gratings to use the instrument as a medium resolution (R~10000) spectrograph. The PIM consist of two flat mirrors and one lens: it allows to image the pupil plane inside the instrument to the detector array and will be used for calibration and alignment purposes. For positioning and locking the optical modules inside the science beam we intend to use an indent mechanism. The MSM work package includes the design, analysis, assembly, alignment and test of the cryogenic mechanism including all optical components and their mounts located on the rotating platform. Cryogenic testing on subcomponent level (motors, indent mechanism) has already started. |
|||||
Impressum --- Datenschutzerklärung Physik Datenschutzerklärung LMU |